![Applied Biosystems 3730xl Скачать руководство пользователя страница 92](http://html.mh-extra.com/html/applied-biosystems/3730xl/3730xl_user-manual_2955116092.webp)
1.
Examine the peaks of the spatial calibration.
A profile is acceptable if the peaks meet the following conditions.
• Peaks in the profile are approximately the same height.
• The peak height increases at the beginning and the end of the spatial profile.
Note:
If you are using a 96-capillary array, a small peak may appear in the left
side of the profile. The peak is normal.
Figure 13
Passing spectral profiles
For additional information, see “Examples of passing and failing spatial
profiles“ on page 94.
A profile is not acceptable if the peak heights are irregular. See “Spatial
calibration troubleshooting“ on page 138.
2.
Verify that an orange cross appears at the top of each peak in the profile.
Expand the profile if needed. See “Magnify the spatial profile“ on page 96.
1
Figure 14
Peak that does not have orange cross
1
No orange cross
Evaluate the
spatial calibration
data
Chapter 8
Calibrate the instrument
Perform spatial calibration
8
92
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide