3.
Place the plate assemblies into the stacker. Ensure that the notched corner of the
plate assembly is at the rear right corner of the stacker.
IMPORTANT!
Do not place more than 16 plates in the stacker.
1
Notched corner
4.
Close the metal In Stack tower door.
5.
Close the Stacker drawer.
After the instrument completes the spectral calibration run, the pass or fail status of
each capillary is listed in the Event Log.
1.
In the navigation pane of the Data Collection Software, select GA
Instruments
4
ga3730
4
Instrument name
4
Instrument Status
4
Event Log.
View the Pass/Fail
status after the
run
Chapter 8
Calibrate the instrument
Perform spectral calibration for sequencing and fragment analysis
8
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
103