Symptom
Possible cause
Action
• Electric discharge
Condensation on the lower part of the
oven door, near the array header.
IMPORTANT!
Ensure that the
environment (humidity) is non-
condensing.
Wipe away spills, moisture, and
condensation with a lint-free lab
cloth. If the problem persists, contact
Thermo Fisher Scientific.
Condensation inside the oven.
There is not enough fluid in larger
chamber of ABC, or the anode buffer
has spilled into smaller overflow
chamber.
Pipette the buffer from the smaller
overflow chamber to the larger
chamber. Ensure that the buffer is
filled to within ±1 mm of the fill line.
When installing new ABC, tilt the
container to move buffer to the larger
side of the container as described in
“Install the anode buffer container
When you remove the heat seal from
a new pouch, some residual seal
remains on top of the pouch.
The top seal of the pouch has become
delaminated and left the polyethylene
behind on the pouch cap.
Use a pipette tip to remove the entire
seal from the pouch cap before
installing on the instrument.
RFID troubleshooting
Symptom
Possible cause
Action
Unable to read RFID information.
“Failure to Read from RFID tag”
Consumable package is improperly
installed or label is defective.
Polymer/Conditioning reagent pouch
is not positioned properly.
Ensure that the RFID label is not
visibly damaged and consumable
package is properly installed.
Ensure that label is close, and
parallel, to the instrument.
Reposition or re-install pouch, then
click Refresh on the Dashboard.
Restart the instrument and the
computer (see “Restart the
instrument and the computer“ on
Appendix A
Troubleshoot
RFID troubleshooting
A
258
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
Содержание 3500
Страница 325: ......
Страница 326: ...thermofisher com support thermofisher com askaquestion thermofisher com 25 October 2018 ...