4.
Specify the starting well for the injection position in which you loaded the
standard in the plate.
Note:
If you navigate away from the Install Standard screen after you start the
install check, the starting well may be reset to A01. This is a display issue only;
the starting well you specify is used for the install check.
5.
Click Start Run.
The instrument performs one run and indicates the number of observed allele and
size standard peaks.
The Capillary Run Data display is updated after the run is complete. The number of
observed size standard and allele peaks is shown. Results for each allele are shown at
the bottom of the screen in the Run Information table.
Number of
peaks per
capillary
Plot and allele
size/height for
selected capillary
Allele results
for all
capillaries
What you see
during a fragment
install check
Chapter 6
Run calibrations and install checks
Run a fragment/HID install check
6
134
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1