Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 4
Data Sheet: Temperature Non-Uniformity Test
Data Sheet: Temperature Non-Uniformity Test
When you run the temperature non-uniformity test, record the setpoint values for the
wells L1 to L6 and R1 to R6 listed on this data sheet. At the end of the test, compare
the values that are displayed on the thermal cycler with the values that are recorded
here.
Note:
You can photocopy this page.
Date
Tested by
Probe serial no.
Thermometer
serial No.
Setpoint value
95 °C
55 °C
L1
L2
L3
L4
L5
L6
R1
R2
R3
R4
R5
R6