Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 4
Temperature non-uniformity test
7. For the GeneAmp
®
9700 with Manual Dual 96-Well Sample Block Module:
Slide the heated cover forward and pull the lever down.
For the GeneAmp
®
9700 with Auto-Lid Dual 96-Well Sample Block Module:
Press the red button to slide and close the heated cover.
IMPORTANT!
Seat the probe properly and the close the heated cover carefully.
If the probe wire is crushed when you close the heated cover, the probe can be
damaged.
8. Power on the digital thermometer.
Refer to the instructions in your Temperature Verification Kit for details on
operating the digital thermometer, Model 4500.
Configuring the
GeneAmp
®
9700
1. Power on the thermal cycler. The Main menu opens.
For best results, allow the thermal cycler 15 to 30 minutes to warm up to
ambient temperature before performing the test.
2. Press
F4
(Util).
The Utilities screen opens.
3. Press
F1
(Diag). The Diagnostics screen opens.
4. Press
F3
(TmpVer). The Temperature Verification screen opens.
5. Press
F2
(TNU). The GeneAmp
®
9700 is automatically configured for the
temperature non-uniformity test, starting with the setpoint of 95 °C, and the
TNU Performance screen opens. The following screen opens when the cover
reaches the setpoint.
Temperature Verification
Temp - Calibration Verification
TNU - Temperature Non-Uniformity
F1
F3
F4
F5
F2
Exit
Temp
TNU
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
F1
F3
F4
F5
F2
Cancel
Place Probe in well L1
Press RUN
RUN