Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 2
Temperature non-uniformity test
Ending the test
1. Press
F5
(Cancel).
2. Remove the 0.2-mL probe assembly from the sample block.
3. Power off the digital thermometer, then clean off the oil.
4. Wait for the sample block to reach room temperature (~25 °C), then remove the
2720 temperature verification frame from the sample block.
PHYSICAL INJURY HAZARD.
Hot Surface. Use care
when working around the heated cover and sample block to avoid being burned
by hot components.
5. Clean the oil off the sample block.