113
Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 9
Temperature non-uniformity test
Calculating test
results
Use the following information to calculate the results of the test.
Note:
If there is more than one GeneAmp
®
9600 in your laboratory, verify that the
serial number on the calibration label matches the serial number on the instrument
that you are testing.
For the 16 Setpoint #1 measurements (95 °C hold), subtract the lowest measured
temperature from the highest measured temperature.
For the 16 Setpoint #2 measurements (40 °C hold), subtract the lowest measured
temperature from the highest measured temperature.
If
either
result is more than
±
1 °C from the setpoint, your GeneAmp
®
9600 must be
serviced by an Applied Biosystems service representative.
Ending the test
1. Remove the 0.2-mL probe assembly from the sample block.
2. Power off the digital thermometer, then clean off the oil.
3. Wait for the sample block to reach room temperature (~25 °C), then remove the
9600 temperature verification frame from the sample block.
PHYSICAL INJURY HAZARD.
Hot Surface. Use care
when working around the heated cover and sample block to avoid being burned
by hot components.
4. Clean the oil off the sample block.