AT851 Battery Aging Tester
4.
Basic Operation
This section includes the following information:
•
CC/CV/CP/CR Operation Mode
•
Setup the Maximum Value
•
Von/off Operation
•
Load ON Timer Function
•
File Operation
•
Menu Operation
4.1
Modes of Operation
The four modes of operation are:
Constant Current (CC)
Constant Voltage (CV)
Constant Power (CP)
Constant Resistance (CR)
4.1.1
Constant Current Mode [CC]
In CC mode, the load will sink a current in accordance with the programmed value
regardless of the input voltage. (See following figure)
Figure 4-1
Constant Current
Mode
Setting
Current
Load
CI)
Load Input V)
AT851 provides two methods to set the current value by using the numeric keypad and
the knob.
Method 1: Setup the current value by using the numeric keypad
Table 4-1 Setup input current (I-Set) by using numeric keypad
Procedure Operation details
Display
Step 1
Press
I-Set
If the mode is not the CC
mode Press
I-Set again
Step 2 Enter a new value using numeric keypad or rotary the knob to adjust the value.