48
AT824/825/826 Handheld LCR Meter
Battery working
time
8h @100% brightness
11h @ 50% brightness:
14h @ 30% brightness
Charge time
About 2h
Lower brightness
5min/10min/20min/30min/OFF
*Power by battery
Auto power off
5min/10min/20min/30min/OFF
*Powered by battery
General
Temperature
0
°
C~40
°
C
Humidity
≤
90%RH
Altitude
2000m
Storage Temperature -10
°
C~70
°
C
Weight
350g
Safety and
electromagnetic
compatibility
IEC 61010-1:2001
IEC 61326-2-1:2005
9.2
Accuracy
Accuracy is defined as meeting all of the following conditions.
Temperature: 23
℃
±5
℃
Humidity:
≤
65% R.H.
Correction: open and short correction
Test terminal: Internal Terminals Slot
Equivalent mode: Auto
Rate: Slow
Warming time: >30 min
A 1-year calibration cycle
Test level accuracy:
10%
Test frequency accuracy
*1
: 0.02%
Basic Accuracy:
0.2%
Tips:
120Hz is nominal frequency, actual frequency is 120.048 Hz,
accuracy is 0.05%
9.2.1
C and D
100Hz/120Hz