27
PLC (Handler) Interface
VCC
(
5V
)
: 34-35
GND:
27-30
E
lectrical Characteristics
Power requirements
:
+3.3V~35VDC
Output signals
:
collector output of built-in pull-up resistor. Darlington drive, low level is
valid.
MAX voltage
:
supply voltage
Input signals
:
opto-isolator. low level is valid.
MAX current
:
50mA
Tip
:
To avoid the damage to interface, supply voltage can not exceed power requirement.
To avoid damage to interface, please connect cable after instrument is powered off.
Output signal can control signal and small pwer relay
(
instrument internal has integrated
reverse energy release diode
)
. Don
’
t use internal power source for large power relay.
Input terminal schematic diagram
Figure 7-2 Input terminal schematic diagram
(
Trig
)
1
2
4
3
U221
P521
R243
499
EXVCC
1
2
3
JP205
GND
1 2 3
JP204
TRIG
In above figure
:
JP204 and JP205 is factory configuration 1-2 short circuit, trigger signal is
rising edge trigger.
If trigger signal is falling edge trigger, please short circuit 2-3 for JP204 and JP205, external
VCC floats.
Output terminal schematic diagram
Figure 7-3 Output terminal schematic diagram
Q1
NPN
Q2
NPN
GND
J1
Internal
J2
TO PLC IO or Signal Relay
R243
2k