![Applent Instruments AT5220 Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/applent-instruments/at5220/at5220_user-manual_2954884025.webp)
25
PLC (Handler) Interface
7.
PLC (Handler) Interface
This chapter describes how to use the handler interface.
Pin Assignment
Connection and Circuit Diagram
By using the handler interface, you can output the measurement completion signal (EOC), the
screening result of the comparator function (GD/NG), and so on to external devices from the
AT5220. You can also input the external trigger signal and the comparator select signal to the
AT5220. With this interface and the comparator function, you can build an automatic
screening system composed of the AT5220 and the handler.
7.1
Pin Assignment
Figure 7-1 Pin assignment
Output signals
(
all signals are low level valid
)
Table 7-1 Input terminal pin definition
Pin
Name
Description
1
CH8-R
0
:
R-OK
2
CH9-R
0
:
R-OK
3
CH10-R
0
:
R-OK
5
NG
All channels are NG output 1
:
OK, 0
:
NG
7
EOC
0
:
in measuring
,
1
:
measuring completed
19
CH7-R
0
:
R-OK
20
CH6-R
0
:
R-OK
21
CH5-R
0
:
R-OK
22
CH4-R
0
:
R-OK
23
CH3-R
0
:
R-OK
24
CH2-R
0
:
R-OK
25
CH1-R
0
:
R-OK
8
CH1-V/RV
0
:
V-OK or RV-OK
Содержание AT5220
Страница 24: ...24 AT5220 User s Guide ...