![Applent Instruments AT5220 Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/applent-instruments/at5220/at5220_user-manual_2954884023.webp)
23
<SYSTEM CONFIG> Page
The configuration of RS-232
The keypad will be locked when the commutation established.
Use SCPI language in Mini-USB programming.
RS-232 configuration is as follows:
•
Data bits: 8
•
Stop bits: 1
•
Parity:
none
•
Baud rate: configurable
Procedure to set baud rate
:
Step 1
Press bottom soft-key [SYSTEM] to enter <SYSTEM CONFIG> page
Step 2
Use cursor keys to select [BAUD] field
Step 3
Use function keys to select
:
Function keys
Function
1200
For communication converter with opt coupler isolation
9600
Communicate with PLC or touch panel, recommend to use this baud rate
38400
Communicate with PLC or touch panel, can use this baud rate
57600
Can use this baud rate when communication with PC
115200
Best for communication with PC
6.1.5
RS-232 Shake Hand [SHAKE HAND]
AT5220 supports software
“
shake hand
”
. AT5220 will return the whole command to host and
then response the command when the [SHAKE HAND] is turned
Procedure to turn ON the
“
Shake Hand
”
:
Step 1
Press bottom soft-key [SYSTEM] to enter <SYSTEM CONFIG> page
Step 2
Use cursor keys to select [SHAKE HAND] field
Step 3
Use function keys to select
:
Function keys
Function
OFF
SHAKE HAND in invalid, no special requirement, please turn off SHAKE
HAND.
ON
6.2
<SYSTEM INFORMATION>
Press
【
Meas
】
or
【
Setup
】
to enter main page, press buttom
【
SYSTEM
】
soft-key to enter
<SYSTEM CONFIGURATION>page
,
then press
【
SYSTEM INFORMATION
】
.
There is no configurable option in the <SYSTEM INFO> page.
Figure 6-2 <SYSTEM INFORMATION>page
Содержание AT5220
Страница 24: ...24 AT5220 User s Guide ...