AT381x User's Guide
78
COMParator
:STATe
:MODE
:TOLerance
:SECONDARY(SLIM)
{ON,OFF}
{ABS,PER,SEQ}
:NOMinal
:BIN
<value>
<low>,<high>
:AUX
{ON,OFF}
:BINS
{1 to 9}
<n,low,high>
:BEEP
{OFF,PASS,FAIL}
:OPEN
{OFF,2,5,10,20,50}
10.12.1
COMParator:STATe
The COMParator:STATe command sets the comparator function to ON or OFF.
Command Syntax
COMParator:STATe {ON,OFF,1,0}
Parameter
Where,
ON or 1
Sets the comparator to ON
OFF or 0
Sets the comparator to OFF
Example
SEND> COMP:STAT OFF
Query Syntax
COMParator:STATe?
Query Response
{on,off}
Example
SEND> COMP:STAT?
RET>
on
10.12.2
COMParator:MODE
The :COMParator:MODE command sets the limit mode of the comparator function.
Command Syntax
COMParator:MODE {ABS,PER,SEQ}
Parameter
Where,{ABS,PER,SEQ} is:
ABS
Absolute tolerance mode
PER
Percent tolerance mode
SEQ
Sequential mode
Example
SEND> COMP:MODE PER
Query Syntax
COMParator:MODE?
Query Response
{abs,per,seq}
Example
SEND> COMP:MODE?
RET>
abs