AT381x User's Guide
56
7.
File Operation
This chapter provides information on the file operation of the AT381x.
The AT381x has built-in non-volatile memory, users can save system configuration data and
user data in this memory. The system's built-in memory can save 10 configuration files.
If you have the USB memory interface option installed, the data can also be saved in an
external USB memory. With USB memory, you can save up to 999 measurement results files in
addition to 10 sets of configuration files.
7.1
<FILE> Page
When press the [Meas] or [Setup] key followed by [FILE] soft key at bottom, the <FILE> page
appears.
Figure 7-1
<FILE> Page
In <FILE> page, users can set the following functions:
[MEDIA] selection – internal memory or external USB memory
[AUTO RECALL] – Recall file 0 or last used file at boot
[AUTO SAVE] on/off – Save the modified data to the current file automatic.
Soft key [SAVE] - Save the current setting data to the current file immediately
Soft key [RECALL] - Load current file data into the system immediately
Soft key [ERASE] – Erase current file data and restore to factory settings.
Soft key [MODIFY DES] – Rename the file description.
7.1.1
[MEDIA] Selection
Select internal memory or external USB storage.
USB memory function cannot be enabled if the USB memory interface is not installed
Procedures for selecting [MEDIA]
Step 1
Press the [Meas] or [Setup] key followed by [FILE] soft key at bottom and then enter <FILE>
page.
Step 2
Use the cursor key to select [MEDIA] field
Step 3
Use the soft keys to select.