45
[Setup] Key
●
Includes this point
○
Excludes this point
In sequential mode, the comparison uses the direct reading measurement value to compare
with the upper and lower limit ranges of the bin. The nominal value does not need to
participate in operation.
To set up the compare mode:
Step 1
Enter the <BIN TABLE> page.
Step 2
Use the cursor key to select [MODE] field
Step 3
Use the soft keys to select comparator mode
Soft key
Function
ABS
Absolute parameter values
PER
Deviation percentages.
SEQ
Sequential mode.
5.3.4
Nominal Value for tolerance Mode [NOM]
You must configure the nominal value when you use tolerance mode as the limit mode for the
primary parameter.
In sequential mode the nominal value does not affect sorting. In sequential mode you can
configure nominal value or not.
NOTE
When using negative nominal values, be sure to set the lower limit to a value higher than the
upper limit because when they are converted to absolute values, the lower limit becomes
higher than the upper limit.
The nominal value entered corresponds to the primary parameter of the test [Function].
To enter the nominal value:
Step 1
Enter the <BIN TABLE> page.
Step 2
Use the cursor key to select [MODE] field
Step 3
Use the numeric keys to enter data, the unit is selected by using soft keys.
5.3.5
Auxiliary Bin ON/OFF [AUX]
If the user does not need to sort the secondary parameters, the auxiliary bin (AUX) can be
turned off.
After shutting it down, the secondary parameter limit will not be determined during the
measurement.
To turn on/off auxiliary bin:
Step 1
Enter the [BIN TABLE] Page
Step 2
Use the cursor key to select [AUX] field