39
[Setup] Key
Step 2.
Use the cursor key to select [MON 1] or [MON 2] field
Step 3.
Use the soft keys to choose a parameter
Soft key
Function
OFF
Turn off the monitor
Z
Absolute value of impedance
D
Dissipation factor
Q
Quality factor(=1/D)
Vac
Test signal Voltage
Iac
Test signal Current
∆
Absolute deviation value
∆
%
Relative deviation value
θ
r
Phase angle (radian)
θ
d
Phase angle
R
Resistance (=Rs)
X
Reactance
G
Conductance
Y
Admittance
5.1.6
Measurement [DELAY]
The instrument can set the delay time before each test by testing [Delay] timer, and wait for
the station to be ready before testing.
The maximum delay time is 60s, the minimum delay time is 1ms.
5.1.7
Auto Level Control [ALC]
The ALC function adjusts the voltage and current across the DUT to match the preset voltage
and current values. A constant level will be obtained on the device under test without being
affected by source internal resistance.
Procedure for setting ALC:
Step 1.
Press the [Setup] key to enter <MEAS SETUP> page
Step 2.
Use the cursor key to select [ALC] field
Step 3.
Use the soft keys to set constant level
Soft key
Function
OFF
Turn off the ALC function.
ON
Turn on the ALC function.
NOTE
When the constant level is turned on, '*' is added at the end of the level setting value,
indicating that the constant level function is turned on.
5.1.8
[NOMINAL] Value Setting
For the convenience of setting, when the [Mon
itor 1] or [Monitor 2] function is set to Δ or Δ%,
the nominal value field is displayed on <Setup> page.
This standard value is synchronized with the standard value save of <BIN TABLE> page.