29
[Meas] Page
ALC ON
Turn Auto Level Control ON
ALC OFF
Turn Auto Level Control OFF
The most commonly used level list
Table 4-11
The most commonly used level list
INCR +/ DECR -
(
V
)
0.01
0.10
0.30
0.50
1.00
1.50
2.00
INCR +/ DECR -
(
A
)
0.0001
0.0005
0.001
0.005
0.01
0.02
4.1.6
Measurement Speed [SPEED]
The instrument offers 3 test speeds (slow, medium and fast). The slower the test speed, the
more accurate and stable the test results are.
Procedure for setting the test speed:
Step 1
Press [Meas] key to enter <MEAS DISPLAY> page;
Step 3
Use the cursor keys to select [SPEED] field;
Step 4
Use soft keys to select
Soft key
Function
SLOW
The measurement period is the longest, about 3 times / sec
MED
Moderate, about 20 times / sec
FAST
The fastest measurement, about 40 times / sec
The measurement speed refers to the time when Handler interface is triggered to the end of
measurement (EOM) output.
1.
Measurement frequency:
300 kHz.
2.
Page:
<BIN MEAS> Page
3.
Range:
Hold range or nominal range
4.
Average:
1
5.
DC Bias:
OFF
6.
Automatic LCZ parameters: OFF
7.
Monitor 1:
OFF
8.
Monitor 2:
OFF
9.
ALC:
OFF
(
ms
)
Test frequency (Hz)
Speed
10
20
100
1k
2k
10k
100k
300k
DCR
SLOW
1600
800
483
342
336
332
332
332
333
MED
1600
800
160
94
91
88.5
88.5
88.5
171
FAST
1600
800
160
30
26.5
24.5
24.5
24.5
48
4.1.7
[LOG] Data
The instrument can record 10,000 sets of test data through the internal data buffer. These data
are saved in an external USB disk in (.csv) file format. These files can be opened on a PC using
a Windows Excel.
Procedure for setting recording data:
Step 1
Press [Meas] key to enter <MEAS DISPLAY> page;
Step 2
Use the cursor keys to select [Record] field;
The option before recording is not enabled: