27
[Meas] Page
Note:
When range is automatic, the instrument will perform range prediction for each measurement
cycle, so test speed will be slightly slower than hold range. Moreover, frequent changes in the
range during automatic measurement can slow down the response. Usually, when instrument
is used as a sorting measurement, auto range method is not suitable.
For the sorting user, please select nominal range mode.
4.1.3
Measurement Frequency [FREQ]
AT3818
:
10Hz
~
300kHz continuous test frequency
AT3816A
:
50Hz
~
200kHz continuous test frequency
AT3816B
:
50Hz
~
200kHz 37 points
AT3817A:
50Hz
~
100kHz continuous test frequency
AT3810A:
10Hz
~
20kHz continuous test frequency
AT3817D: 50Hz
~
100kHz, 10 points
Table 4-6
AT381x frequency resolution
Frequency range(F)
Resolution
10.0000Hz
≤
F
≤
99.9999Hz
0.0001Hz
100.0000Hz
≤
F
≤
999.999Hz
0.001Hz
1.00000kHz
≤
F
≤
9.99999kHz
0.01Hz
10.0000kHz
≤
F
≤
99.9999kHz
0.1Hz
100.000kHz
≤
F
≤
300.000kHz
1Hz
Frequency accuracy
:
0.01%
Procedure for setting the test frequency
:
Step 1
Use cursor keys to select [Frequency] field;
Step 2
Use cursor keys to select [Frequency] field;
Soft key
Function
INCR +
Frequently used frequency list
DECR -
Frequently used frequency list
The most commonly used frequency list
Table 4-7
AT3818 the most commonly used frequency list
INCR +/ DECR -
10Hz
50Hz
60Hz
100Hz
120Hz
1kHz
10kHz
20kHz
40kHz
50kHz
100kHz
200kHz
250kHz
300kHz
Table 4-8
AT3816A/AT3816B List of the most common frequencies
INCR +/ DECR -
10Hz
50Hz
60Hz
100Hz
120Hz
1kHz
10kHz
20kHz
40kHz
50kHz
100kHz
200kHz
Table 4-9
AT3817A Frequency list
INCR +/ DECR -
10Hz
50Hz
60Hz
100Hz
120Hz
1kHz
10kHz
20kHz
40kHz
50kHz
100kHz
Table 4-10
AT3810A List of the most common frequencies
INCR +/ DECR -