AT381x User's Guide
16
2.
Overview
This chapter contains general information about AT381x. The information is organized as
follows
Introduction
Measurement Function
Signal Source
Main Functions
2.1
Introduction
Thank you for purchasing AT381X LCR meter.
AT381X is precision LCR meter that uses a fully automated real-time inspection micro-desktop
instrument controlled by high-performance 32-bit ARM microprocessor. The instrument can
select any test frequency between 10Hz and 300 kHz, and can select test signal level between
0.01V and 2.00V with 0.01V steps. Built-in -2.5V~+2.5V programmable DC bias automatically
measures inductance L, capacitance C, resistance value R, complex impedance Z, quality factor
Q, loss tangent value D, phase angle
θ
(degrees), phase
θ
(radian), and DC resistance DCR.
The primary and secondary parameters are displayed in all six digits; there are two monitoring
displays at the same time. One of Z, D, Q,
θ
r,
θ
d, R, X, G, B, Y, Vac, Iac,
∆
and
∆
% can be
displayed at the same time. With an accuracy of up to 0.05%, the instrument can meet
requirements of accurate testing and mass production by various component manufacturers,
schools, research institutes and metrological quality inspection departments.
The instrument has a professional sorting function, with 10 sets of storage files,
programmable 9 qualified file, 1 auxiliary file (secondary-parameter unqualified), 1
unqualified file and primary parameter HI/IN/LO file, can set percentage points or absolute
value sorting, equipped with Handler interface and RS-232C interface, used in automatic
sorting system to complete automatic pipeline testing. An optional USB memory interface
allows users to save setup data and measurement data to an external mover.
The instrument supports two kinds of communication protocols: computer remote control
instruction is compatible with SCPI (Standard Command for Programmable Instrument
standard command set) and Modbus communication protocol, remote control and data
acquisition functions are efficiently completed.
Reference
:
See 11. Specifications section for complete technical specifications.