AT381x User's Guide
116
01
03
31 14
3A 83 12 6F
3C 23 D7 0A
51 61
Lower limit
Upper limit
12.4.8
Primary Parameter Limit Value Register [3110H-3133H]
There are total of 9 bins for primary parameter, the limit value starts from 3110, the lower limit
of each bin uses 2 registers, and the upper limit uses 2 registers, total of 4 registers.
The lower and upper limits can be set separately, or simultaneously.
Write
BIN1
Lower limit
:
-10, Upper limit
:
10
1
2
3~4
5
6
7
8~11
12~15
16~17
01
10
3110 00
02
04
C1 20 00 00
41 20 00 00
CD5C
Lower limit
Upper limit
Response:
1
2
3
4
5
6
7
8
01
10
31
10
00
04
8F
1F
Read
1
2
3
4
5
6
7
8
01
03
31
10
00
04
4B
30
Response:
1
2
3~4
5~8
9~12
13~14
01
03
31 10
C1 20 00 00
41 20 00 00
6C7F
Lower limit
Upper limit
12.5
File Operation
Since the AT381x settings are stored in the file, after the Modbus command is set, the data
cannot be stored in the internal FlashRom in real time, which will cause the register data
before the next power-on to be restored to the original file value.
Users can store all set values in the current or specified file with the file manipulation registers.
Meanwhile, the specified file data can also be recalled into the setup register.
12.5.1
Save to Current File [4000H]
Send a value of 0001 to 4000 registers, the AT381x will perform a file write operation, and all
settings will be saved to the current file.
This register cannot be read.
Write
1
2
3
4
5
6
7
8
9
10
11
01
10
40
00
00
01
02
00
01
26
54
write
register
number of
registers
byte
data
CRC
Response:
1
2
3
4
5
6
7
8
01
10
40
00
00
01
14
09