115
Modbus
(
RTU
)
Command
0001
PASS
OK beep
0002
FAIL
NG beep
12.4.6
Nominal Value Register [310AH-310BH]
The primary parameter nominal value uses 2 registers, 310A and 310B. Please note that
reading 310B alone is invalid.
Write
100E-9 (Single precision floating point number
:
33D6BF95)
1
2
3
4
5
6
7
8
9
10
11
12
13
01
10
31
0A
00
02
04
33
D6
BF
95
74
A2
write
register
number of
registers
byte
data
CRC
Response:
1
2
3
4
5
6
7
8
01
10
31
0A
00
02
6F
36
register
number of
registers
CRC
Read
1
2
3
4
5
6
7
8
01
03
31
0A
00
02
EA
F5
read
register
number of
registers
CRC
Response:
1
2
3
4
5
6
7
8
9
01
03
04
33
D6
BF
95
A4
D0
byte
data 100E-9
CRC
12.4.7
Secondary Parameter Limit Value Register [310CH-310FH]
The secondary parameter limit value starts from 310C, the lower limit uses 2 registers, and the
upper limit uses 2 registers, total of 4 registers.
The lower and upper limits can be set separately, or simultaneously.
Write
Lower limit
:
0.001, Upper limit
:
0.01
1
2
3~4
5
6
7
8~11
12~15
16~17
01
10
310C 00
02
04
3A 83 12 6F
3C 23 D7 0A
21 AE
Lower limit
Upper limit
Response:
1
2
3
4
5
6
7
8
01
10
31
0C
00
04
0F
35
read
1
2
3
4
5
6
7
8
01
03
31
0C
00
04
8A
F6
Response:
1
2
3~4
5~8
9~12
13~14