AT381x User's Guide
108
01
03
04
3F 80 00 00
F7
CF
byte
data
CRC
Note: The following situations will return an error
Test Condition
Description
<CORRECTION>
page
Error in operation
<List Sweep> page
Error in operation
Level is current state
Error in operation
Function
:
DCR
Error in operation
12.3.9
DCR Range Register [300AH]
Write
1
2
3
4
5
6
7
8
9
10
11
01
10
30
0A
00
01
02
00
01
97
3A
write
register
number of
registers
byte
data
CRC
Response:
1
2
3
4
5
6
7
8
01
10
30
0A
00
01
2E
CB
register
number of
registers
CRC
Read
1
2
3
4
5
6
7
8
01
03
30
0A
00
01
AB
08
read
register
number of
registers
CRC
Response:
1
2
3
4
5
6
7
01
03
02
00
04
B9
87
byte
data
CRC
Data values
:
Data
Function
Description
0000
Range 0
100k
Ω
0001
Range 1
30k
Ω
0002
Range 2
10k
Ω
0003
Range 3
3k
Ω
0004
Range 4
1k
Ω
0005
Range 5
300
Ω
0006
Range 6
100
Ω
0007
Range 7
30
Ω
0008
Range 8
3
Ω
12.3.10
Startup File Recall Register [300CH]
Write