background image

p  //o

Healthcare 

Technologies 

APH183

™ 

Apollo 5 

Premium

USER MANUAL 

www.apollo-ht.co.uk 

AHSM-005_Version_3_21.10.2021

Содержание 5 Premium

Страница 1: ...p oHealthcare Technologies APH183 Apollo 5 Premium USER MANUAL www apollo ht co uk AHSM 005_Version_3_21 10 2021...

Страница 2: ...TABLE OF CONTENTS PAGE INSTRUCTIONS 2 INTRODUCTION 4 PRODUCT DESCRIPTION 5 INSTALLATION 7 OPERATION 8 CLEANING 10 STORAGE 10 MAINTENANCE 10 SPECIFICATIONS 12 SAVE THESE INSTRUCTIONS 13...

Страница 3: ...n or repair 3 Do not drop or insert any object into any opening or hose 4 Connect the unit to a properly grounded outlet only See grounding instruction 5 When plugged in the unit should not be left un...

Страница 4: ...ron Tumble Dry Normal Low Heat Do Not Tumble Dry Do Not Bleach Do Not Dry Clean Machine wash regular normal max 95 degrees C 203 degrees F Attention Observe proper Disposal of Electrical Electronic Eq...

Страница 5: ...11 IEC EN 61000 3 2 Class A IEC EN 61000 3 3 CISPR 11 Group 1 Class B A 2 INTENDED USE The unit should help prevent and reduce the incidence of pressure ulcers while optimising Patient com fort It is...

Страница 6: ...Punpurnt 0 ttress System 0Olli C llYreC ll lfS 8 2 PUMP AND MATERIALS 0 Comfort Adjust button 30 150kg f Power On Off button E Low Pressure Indicator 0 Normal Pressure Indicator 8 Panel Tactile 8 Fus...

Страница 7: ...the top of the switch The switch will now be in the Off position 3 When the pump is first turned on the low pressure light illuminates whilst the mattress is inflating Once the mattress has reached t...

Страница 8: ...re mains wire could be a risk on the floor under the bed with the use of other equipment clip the mains cable into the cable management system on the side of the mattress 4 Connect the air hose connec...

Страница 9: ...in alternating mode There i s no stati c mode on thi s product 4 Adjust the comfort setting to the most comfortable level for the Patient without the mattress bottoming out The pressure in the mattre...

Страница 10: ...D will light D 4 LOW PRESSURE INDICATION FUNCTION Once the mattress has reached the desired comfort setting should the pump detect low pressure in the mattress the Low Pressure LED will light up Orang...

Страница 11: ...so be washed in the washing machine at a temperature not exceeding 95 C See printed washing instructions on the cover DO NOT USE PHENOLS Always dry the Mattress without direct exposure to sunlight F S...

Страница 12: ...rewdriver 3 Insert a new fuse of the correct rating and replace the cover The fuse should be rated as T1A 250V and CE approved G 3 AIR FILTER REPLACEMENT 1 Remove the air filter cover located at the b...

Страница 13: ...t Operation 10 o 90 non conde sing orage 10 to 90 non condensing Sh pping 10 o 90 noll Cond sing Humidity Class ica io Class 11 ype BF IPX1 Applied Part Mattress DO NOT USE IN THE PRESENCE OF A FLAMMA...

Страница 14: ...non condensing Classification Class II Type BF IP21 Applied Part Air Mattress Not suitable for use in the presence of a flammable anesthetic mixture No AP or APG protection Mattress Specification Mod...

Страница 15: ...lies buildings used for domestic purposes Harmonic emissions lEC 61000 3 2 Class A Voltage fluctuations flicker emissions lEC 61000 3 3 Complies Warning 1 The device should not be used adjacent to or...

Страница 16: ...ations equipment should be used no closer to any part of the models P08 including cables than there commended separation distance calculated from the equation applicable to the frequency of the transm...

Страница 17: ...ation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2xP1 2 80MHz to 800MHz d 1 2xP1 2 800MHz to 2 5GHz d 2 3xP1 2 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8...

Страница 18: ...epair or replace the unit unless this is by qualified personal To reduce the risk of shock or electrocution do not change or modify the cord or plug in any way oHealthcare Technologies For further tec...

Страница 19: ...co uk www apollo ht co uk Due to ongoing research and development Apollo Healthcare Technologies Ltd reserve the right to change specifications without prior notice This will not affect the efficacy...

Отзывы: