MQ series ePAR meters are designed with a user-friendly interface allowing quick and easy measurements.
Press the power button to activate the LCD display. After two minutes of non-activity the meter will revert
to sleep mode and the display will shut off to conserve battery life.
Press the mode button to access the main menu, where manual or automatic logging are selected, and
where the meter can be reset.
Press the sample button to log a reading while taking manual measurements.
Press the up button to make selections in the main menu. This button is also used to view and scroll
through the logged measurements on the LCD display.
Press the down button to make selections in the main menu. This button is also used to view and scroll
through the logged measurements on the LCD display.
Logging:
To choose between manual or automatic logging, push the mode button once and use the up/down
buttons to make the appropriate selection (SMPL or LOG). Once the desired mode is blinking, press the mode
button two more times to exit the menu. When in SMPL mode press the sample button to record up to 99 manual
measurements (a counter in the upper right-hand corner of the LCD display indicates the total number of saved
measurements). When in LOG mode the meter will power on/off to make a measurement every 30 seconds. Every
30 minutes the meter will average the sixty 30 second measurements and record the averaged value to memory.
The meter can store up to 99 averages and will start to overwrite the oldest measurement once there are 99
measurements. Every 48 averaged measurements (making a 24-hour period), the meter will also store an
integrated daily total in moles per meter squared per day (mol m
-2
d
-1
).
Reset:
To reset the meter, in either SMPL or LOG mode, push the mode button three times (RUN should be
blinking), then while pressing the down button, press the mode button once. This will erase all the saved
measurements in memory, but only for the selected mode. That is, performing a reset when in SMPL mode will
only erase the manual measurements and performing a reset when in LOG mode will only erase the automatic
measurements.
The LCD display consists of the total number of logged
measurements in the upper right-hand corner, the real-time PFD
value in the center, and the selected menu options along the
bottom.