YD300 Portable Water Hardness Meter   

Quick Start Manual 

 

Scan to watch the video tutorial 

 

 

1. Electrode Activation 

a)

 

For first-time use — shake electrode with force to let the inner solution of electrode fall into the measuring head 

compartment; there should NOT be any bubbles within the measuring head. Pour a small amount of soaking 

solution into the soaking solution vial, and soak the electrode for 30 minutes. 

b)

 

Regular Usage — The activation time only takes 10 to 15 minutes for a frequently used electrode. As long as the 

reading is stable, users can start to perform calibration and measurements (please note that the readings without 

electrode being immersed in solutions are random). 

 

2. Calibration (take B2, B3 calibration and mmol/L as an example) 

1)

 

Power  on  the  meter.  Press                ,                and          icons  appear  and  are  flashing  at  the  same  time, 

indicating the meter enters 1st point calibration; the calibration solution is B2. 

2)

 

Rinse  the  electrode  in  distilled/deionized  water  and  shake  off  excess  water;  insert  the  electrode  into  B2 

calibration solution and gently stir the electrode for a few seconds and let it stand still; Wait for the reading to 

fully stabilize (about 2 minutes). 

3)

 

Then  press                to  confirm  calibration,  “

2.00×10-1mmol/L

”  icon  starts  flashing  (see  picture  (4-3), 

indicating 1st point calibration will be finished after approximately 20 seconds. During the icon flashing, do 

NOT press any key.   

4)

 

NOTE:

 The reading may be temporarily stable in about 30 seconds, but we recommend waiting about 2 minutes 

for  the  reading  to  get  fully  stabilized  before  pressing              to  confirm  calibration.  This  practice  will 

significantly improve the accuracy of calibration. 

5)

 

Then                  and            icons will appear and start flashing, indicating the meter enters 2nd point calibration 

mode; the calibration solution is B3. 