  The silicone oil applied should not be too much. After wiping with micro-fiber cloth, please wipe 

with filter paper or high-quality tissue paper to clean off. Excessive residual silicone oil on the 

vial surface will affect the measurement accuracy. 

  It's not necessary to use silicone oil for each calibration and measurement. Apply silicone oil 

every several days or once a week. In between, just clean the surface with filter paper or high-

quality tissue paper. 

  Clean the calibration vial and sample vial together and keep the steps and actions consistent to 

achieve same degree of cleanliness. 

  Stability of calibration solutions 

The U.S EPA approved AMCO polymer standard calibration solution is very uniform and stable. 

It does not precipitate, drift or condense. It can be used directly without shaking or flipping the 

vial (to make the solution even). For polymer calibration solutions that have not been used for a 

long time, slowly flip the vial twice and  let it stand for 2 minutes. Be careful not to shake the 

solution vigorously, as this will create air bubbles which destabilize the measurement; For 0.0 

NTU calibration solution, do NOT shake or flip. If using Formazin calibration solution, as it tends 

to precipitate easily, each time users must flip and shake the vial to make the solution even. But 

sediment  can  still  occur  during  tests  and  make  the  measurement  unstable.  Operating 

experience is important in using Formazin calibration solutions. 

 

4.2 

Calibration Procedure (Take 0 NTU and 20 NTU as an example) 

1) Instrument warm up: Power on the instrument, and then wait for 3 minutes.

 

2) Open the flip cover and place the 0.0 NTU calibration 

vial. Align the arrow on the vial lid with the one on the 

sample vial holder and close the flip cover.

 

   

3) Press          to enter the calibration menu, the cursor is 

at 0 NTU. Press                  to start calibration.

 

 

 

Содержание TN400

Страница 1: ...1 TN400 Portable Turbidity Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com...

Страница 2: ...Display Mode 6 3 4 Keypad 7 3 5 Power Supply 7 4 Meter Calibration 8 4 1 Preparation for Calibration 8 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 9 4 3 Notes for Calibration 10 5 T...

Страница 3: ...characteristics Smart functions such as automatic calibration operating instructions automatic power off low voltage display etc Large TFT color screen with blue background in measuring mode green bac...

Страница 4: ...s AMCO Polymer or Formazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measurement Mode Normal measurement and average measurement mode Display TFT color screen Sample Vials 25...

Страница 5: ...vial holder Close the cover when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark...

Страница 6: ...0 NTU Carrying case TN400 Turbidimeter Microfiber cloth Screwdriver to remove battery cover Silicone oil 10 ml AA Alkaline battery 4 Sample vial 6 Instruction manual Diagram 2 Measurement mode Measure...

Страница 7: ...erform single shot measurement long press the key to perform continuous measurement In calibration mode press the key to confirm the calibration In calibration mode press the key to select standard so...

Страница 8: ...your supplier to replace See the detail in Section 7 Replacement Parts b Replace zero turbidity solution Replace with 0 0 NTU calibration solution open the vial cap pour out the original solution add...

Страница 9: ...the vial to make the solution even For polymer calibration solutions that have not been used for a long time slowly flip the vial twice and let it stand for 2 minutes Be careful not to shake the solu...

Страница 10: ...00 NTU calibration vial in the sample vial holder and press to calibrate After calibration is done press to confirm repeat the steps to calibrate 400 NTU and 800 NTU After calibration is done press to...

Страница 11: ...ation solution If the wrong solution is selected for calibration the measurement will be completely wrong To fix this problem simply re calibrate the meter selecting the correct calibration solutions...

Страница 12: ...s of continuous measurement displaying 1 data every 8 seconds and will display the average value at the end see diagram 6 The average measurement mode can be used for observing the stabilizing process...

Страница 13: ...d for measurement Please do not wash the sample vial holder as it may damage the optical structure 6 WARRANTY We warrant this instrument to be free from defects in material and workmanship and agree t...

Страница 14: ...L 1 bottle T200 1 20 100 400 800 NTU AMCO polymer Compatible with TN400 TN480 4 bottles T500 3 Sample vials with lid 25 60mm 1 6 6 pcs TN400 S3 Silicone oil 10mL 1 bottle APERA INSTRUMENTS LLC Address...

Отзывы: