1. Brief Introduction 
 

Thank you for purchasing SX630 ORP Pen Tester. This tester is featured with reliable measuring 

performance with a single-button operation in a slim body. The ORP probe is replaceable and suitable for 

test tube measuring and other small-volume sample testing. 

 

2. Specifications 

1) 

Range: ±1999 mV     

2) 

Resolution: ±1 mV 

3) 

Accuracy: ± 0.2% F.S 

4) 

Power: 2 x CR2032 Cell Batteries 

5) 

Dimension and Weight: 148*29*14mm/43g 

6) 

Waterproof rating: IP57 

 

3. Instructions 

 

1) 

Measurement: Rinse the probe in distilled water and dry it. Dip the probe in the sample solution. Stir 

gently and wait until you receive a stable reading.

 

2) 

 After measurement, rinse the probe with distilled water, dry it, and put the protection cap back on. It’s a 

good practice to pour a bit storage solution that comes with kit into the protection cap when storing the 

meter, since it will help keep the accuracy of the probe.

 

4. Notes 
 

1) 

It is no need to calibrate when testing ORP, but if suspect to the performance of meter or the 

measurement result, then can use ORP standard solution to test to identity whether the ORP electrode 

or meter is accurate. The meter comes with the following standard solution: 220mV ± 15mV (25°C). 

2) 

After the electrode has been used over a long period of time, the platinum surface will get polluted 

which causes inaccurate measurement and slow response. Please refer to the following methods to 

clean and activate ORP electrode:  

(a) For inorganic pollutant, soak the electrode in 0.1mol/L dilute hydrochloric acid for 30 minutes, then 

wash it in distilled water, then soak it in the 3M KCL storage solution for 6 hours. 

(b) For organic or lipid pollutant, clean the platinum surface with detergent, then wash it in distilled 

water, then soak it in the 3M KCL storage solution for 6 hours. 

(c) For heavily polluted platinum surface on which there is oxidation film, polish the platinum surface 

with toothpaste, then wash it in distilled water, then soak it in the 3M KCL storage solution for 6 hours. 

3) 

The ORP electrode of this meter can be replaced. The replacement probe’s model number is SX635. 

Please go to www.aperainst.de to watch a tutorial video for how to replace the probe. 