SX610 pH Pen Tester Kit User Manual 

1. Brief Introduction 

Thank  you  for  purchasing  APERA  INSTRUMENTS  SX610  pH  Pen  Tester.  This  product  is  featured  with 
reliable  and  accurate  measuring  performance  with  a  single-button  operation  in  a  slim  body,  ideal  for 
measuring  pH  in  general  water  solution  applications  such  as  hydroponics,  greenhouse,  aquaculture, 
aquarium,  water  treatment,  pools&  spas,  cooling  towers,  beverage  making,  education,  environmental 
monitoring, body fluid (e.g. urine, saliva). The pH probe is replaceable and suitable for test tube measuring 
and other small-volume sample testing. 

2. Specifications 

1)

 

Range: 0 to 14.0 pH; 32 to 140°F (0 to 60°C)

 

2)

 

Resolution: 0.1 pH; 0.1°C

 

3)

 

Accuracy: ± 0.1 pH ±1 digit; ±1°C

 

4)

 

Automatic temperature compensation: 32 to 140°F (0 to 60°C)

 

5)

 

Automatic calibration: 1-point calibration (pH 7.0)

 

6)

 

Power: CR2032 Cell Batteries

 

7)

 

Dimension and Weight: 148*29*14mm/49g

 

8)

 

Waterproof rating: IP57

 

3. Instructions 

1)

 

Calibration: Rinse the probe in distilled water and shake off excess water. Immerse the probe in 
the standard buffer solution pH7.00. Stir for a few seconds and wait for the reading to stabilize. 
Then  press  the  CAL  button  for  about  two  seconds  until 

CAL

  appears  on  the  screen.  When  the 

calibration value 

7.0

 flashes, the calibration will finish after several seconds, and the meter will 

return to measurement mode and the icon        will appear on the screen.

 

2)

 

Measurement: Rinse the probe in distilled water and shake off excess water. Immerse the probe 
in your sample solution. Stir the probe for a few seconds and wait for the reading to stabilize, then 
record the measurement.

 

3)

 

After  measurement,  rinse  the  probe  with  distilled  water,  shake  off  excess  water,  and  put  the 
protection cap back on.

 

4)

 

Storage: 

Always

 leave a bit 3M KCL storage solution in the probe cap when storing the meter as it 

will optimize the performance and maximize the life of the pH probe.

 

5)

 

Change temperature unit: When meter is turned off, hold the CAL button until the meter reboots 

and then the temperature unit will be switched (˚C–˚F)

 

4. Notes 

1)

 

Calibrate the pH pen at least once a month.

 

2)

 

To ensure the accuracy, make sure the pH standard buffer is fresh and clean. The small bottle of 