Apera Instruments PHB-3 Скачать руководство пользователя страница 1

 

 

PHB-3 Pocket pH Tester Instruction Manual 

 

1. Brief Introduction 
 

Thank you for purchasing APERA INSTRUMENTS’ PHB-3 Pocket pH Tester. This tester is simply 
the most economical choice of a reliable pocket pH tester for your regular liquid's pH 
measurement, including hydroponics, aquaculture, pools & spas, home brewing, education, 
tap/drinking water, etc. 
 

2. Specifications 

1)

  pH Range: 0 ~ 14.0 pH     

2)  

Resolution: 0.1 pH 

3)  

Accuracy: ± 0.1 pH, ±1 digit 

4)  

Automatic calibration: 1-point calibration (pH 7.00) 

5)

  Power

2 AAA Batteries, for 1000 hours of use 

6)

  Dimension and Weight

39×25×176 mm / 87 g 

7)

  Waterproof rating: IP57 

 

3. Instructions 
 

1)

 Calibration: Rinse the probe in distilled water and dry it. Dip the probe in the standard buffer 

solution pH7.00. Stir gently, hold still, and wait for the reading to become stable. Then press the 
CAL button for two seconds until 

CAL

 appears on the screen. When the calibration value 

7.0

 

flashes, the calibration is complete and 

END

 will appear on the screen. 

2)

 Measurement: Rinse the probe in distilled water and dry it. Dip the probe in the sample 

solution. Stir gently and wait until you receive a stable reading.  

3)

 After measurement, rinse the probe with distilled water, dry it, and put the protection cap back 

on. 

 

4. Notes 
 

1)

 The instrument can be continuously used for 1 to 2 weeks (or longer) after calibration. 

2)

 The sponge in the protection cap should always be kept wet by adding a few drops of distilled 

water in order to keep the probe activated. 

3)

 To ensure the reliability of calibration, pH standard buffer solution should be replaced once it 

becomes contaminated. 

4)

 Maintenance is necessary for the head of the probe because it can become contaminated 

after long-term use. To do the maintenance, wash the bulb and surrounding area with a small soft 
brush and soap water. Then rinse the bulb with distilled water, and redo calibration after several 
hours.  

5)

 The instrument will be power off automatically without operation after 10 minutes. 

6)

 Replace the batteries when          appears on screen. 

 
 
 
 