LabSen 211 Routine pH Electrode 

User Manual 

LabSen electrochemical sensors are premium pH electrode with manufacturing technology and key 
components imported from Switzerland. LabSen 211 routine pH electrode is suitable for routine use, 
especially suitable for high-precision pH measurement of scientific research and quality control.   
This probe has following features: 

  Impact-resist  membrane  (see  the  right  picture)

,  there  is  no  danger  of  electrode 

breakage during normal use. 

 Blue gel inner solution, does not flow and will not cause bubble. 

 Long life reference system, has better stability and service life. 

 

Technical Data 

Measuring Range

 

(0-14) pH 

Electrolyte

 

3M KCl 

Temperature Range

 

(-5~100)°C

 

Soaking Solution

 

3M KCl 

Shaft Material

 

Lead-free Glass

 

Electrode Dimension

 

(Ø12×120) mm 

Membrane Shape

 

Half Ball 

Connector

 

BNC

 

Reference 

Long Life 

Cable 

Ø3×1m

 

Junction 

Ceramic 

 

 

 

Usage and Maintenance 

2.1 

Connect BNC plug to the input on the pH meter. 

2.2 

When  measuring,  please  unscrew  the  sleeve  cap,  pull  out  the  electrode  and  rinse  it  with 
deionized or distilled water. After using, please put the electrode back into the bottle and screw 
tight the cap.   

 

 

 

 

 

 

 

 

 

 

 

 

2.3 

Prior to measurement, remove the rubber plug to maintain pressure of the reference solution, 
keep consistent flow rate of reference solution and stable potentials of junction. 

2.4 

After a period of usage, the reference solution will running low. Whenever the level falls to 1/2 
height of the electrode, add 3M KCL solution to the refilling hole by using syringe or pipette. 

2.5 

The  connector  of  the  electrode  should  be  kept  clean  and  dry.  If  being  contaminated,  please 
clean it with medical cotton and absolute alcohol and blow dry to prevent the short circuit of the 
electrode and slow reaction of electrode. 

Before  pulling  out  or  putting  back  the  electrode,  make 
sure to fully loosen the blue cap on the storage sleeve so 
that the electrode can move in and out smoothly. 