background image

 

- 13 - 

the probe. Tighten the locking cap. Be careful not to have water droplets on the head of 
the  sensor  cap.  Wait  for  5  to  10  minutes  after  turning  on  the  instrument  in  order  to 
saturate  the  air  in  the  calibration  sleeve  with  water  vapor.  In  addition,  wait  for  the 
temperature to completely stablize.   

(C) 

Long press          to enter the calibration mode, and 

CAL 

is flashing in the upper right 

corner. Wait for  the  stable        to  appear  and  stay  on,  press 

       

to  finish  calibration, 

once 

100%

 starts flashing, the saturated oxygen calibration is completed. 

 

6.2 

Zero-Oxygen Calibration      

Zero-Oxygen  calibration  is  only  performed  when  a  probe  or  sensor  cap  is  replaced,  the 
probe has not been used for a long time, or users have a requirement for high accuracy in 
low DO range (0 to 2.0 ppm). Zero-Oxygen calibration is done at the factory before shipment, 
so it is not necessary to perform it during initial use. Zero-oxygen calibration should follow 
these steps:   
(A)  Preparation of 100ml of oxygen-free water: in the 100ml beaker, weigh 2g of 

anhydrous sodium sulfite (Na

2

SO

3

) and add 100ml of distilled or deionized water to 

dissolve. Oxygen-free water is only effective within 1 hour.   

(B)  Put the electrode into the oxygen-free water, wait for 3 to 5 minutes after the 

instrument is turned on, and wait for the temperature and DO reading to completely 

stabilize. The DO reading should be very close to 0, 0.1mg/L (ppm) or so.     

(C)  Long press          to enter the calibration mode. 

CAL

 will be flashing at the upper right 

corner. Wait for a stable        . Press          and the zero-oxygen calibration is completed. 
 

6.3 

Special Notes for Calibration     

(A) 

To ensure the optimal measuring accuracy, please perform the calibration at the same 
temperature as your samples. It is recommended that the oxygen saturation calibration 
be performed according to section 6.1 before use every day. 

(B) 

A  dried  out  surface  coating  of  the  sensor  cap  can  adversely  affect  the  stability  of 
measurement.  Please  pay  special  attention  to  this  situation.  See  Section  4.2  (Probe 
Maintenance) for details.   

(C) 

The instrument has factory default setting function, select YES in parameter setting P3.6, 
the meter will be calibrated to the theoratical value. 
 

Measurement 

 

7.1 When measuring, place the probe in your sample solution, stir quickly for a few 

seconds in the solution to remove bubbles from the measuring surface of the sensor cap. 

Then hold the probe still and wait for a stable measurement. The solution must be above 

the immersion line of the probe.   

Содержание DO8500

Страница 1: ...1 DO8500 Portable Optical Dissolved Oxygen Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com v1 2...

Страница 2: ...3 Sensor Cap 9 5 Preparation for Calibration 10 5 1 Dissolved Oxygen Units Selection 10 5 2 Resolution Selection 11 5 3 Temperature Unit Selection 11 5 4 Air Pressure Compensation 11 5 5 Salinity Comp...

Страница 3: ...up to 8000 hours of service life Easy to calibrate and maintain Probe is equipped with a calibration storage sleeve which makes calibration and maintenance more convenient and reliable 1 2 Intelligent...

Страница 4: ...tion Points Saturation Point Zero Oxygen Temperature Compensation Automatic 0 to 50 Air Pressure Compensation Automatic 60 to 120 kPa Salinity Compensation Manual or automatic 0 to 45 ppt Temperature...

Страница 5: ...called Timing storage Temperature value and prompts of special display mode Date time and prompts of special display mode Automatic reading lock up Units of date and time ATC Auto temperature compensa...

Страница 6: ...r 2 seconds to turn off Short Press In measurement mode press the key to switch unit mg L or ppm Long Press In the measurement mode press the key for 2 seconds to enter the calibration mode Short Pres...

Страница 7: ...ll be displayed on the screen as shown in Figure 5 DO value 8 18mg L temp value 25 2 When salinity probe is connected and salinity is greater than 1ppt salinity will be displayed on the bottom right o...

Страница 8: ...ashing and the meter stops the auto timing storage In auto timing mode manual storage does not work Set time to 00 00 in parameter P4 1 to exit from the auto timing storage mode 3 6 3 Recall stored va...

Страница 9: ...hould be above this line 4 2 Probe Maintenance The sensor cap of the optical DO electrode must be kept in a moist environment If the surface coating of the sensor cap dries out the reading will be uns...

Страница 10: ...ated please do not use alcohol or organic solvents to clean otherwise it may damage the probe It can be gently wiped with a soft clean cloth To disinfect the probe immerse it in 3 hydrogen peroxide fo...

Страница 11: ...n general users do not have to calibrate air pressure any more If necessary calibrate it according to standard value measured by aneroid barometer Refer to parameter setting P3 5 for the procedure of...

Страница 12: ...e refer to Figure 13 Press to return to measurement mode Special notes A Use 12 88mS cm conductivity calibration solution to calibrate salinity If the standard solution is wrong Err icon will be flash...

Страница 13: ...ffective within 1 hour B Put the electrode into the oxygen free water wait for 3 to 5 minutes after the instrument is turned on and wait for the temperature and DO reading to completely stabilize The...

Страница 14: ...influenced by temperature and the probe senses temperature slower than dissolved oxygen 8 Parameter Settings Press in the measurement mode to enter the parameter setting mode P3 0 press to switch the...

Страница 15: ...1 Press Aut flashes Press to select auto salinity compensation Aut or manual salinity compensation Hnd Press to confirm 2 Press to enter P3 4 mode or press to return to measurement mode P3 4 Salinity...

Страница 16: ...to measurement mode P4 2 temperature unit 1 Press flashes Press to select Press to confirm 2 Press to enter P4 3 mode or to return to measurement P4 3 Auto Lock Off On 1 Press Off flashes Press to se...

Страница 17: ...es press or to adjust value then press to confirm 2 Press to enter in P4 8 mode or press to return to measurement P4 8 Adjust time 1 Press HH flashes then press MM flashes When the number flashes pres...

Страница 18: ...nload press this key to download the data from the meter to the computer Clear press this key to clear all the stored values Export press this key to export the stored value to a Microsoft Excel file...

Страница 19: ...e measuring information is downloaded to the computer through USB and will not be stored in the meter b The stored mode and unit during operation is the same as the data shown on the meter press to ch...

Страница 20: ...the warranty A Fails to install operate or use the product in accordance with the instruction manual or if the product is damaged by abuse or incorrect use B The sensor cap is damaged by external forc...

Страница 21: ...t can be replaced DO readings are not accurate 1 Check whether the temperature is stable the salinity input and barometric pressure are accurate 2 If the probe calibration is not good recalibrate 3 Ch...

Страница 22: ...0 9 67 9 15 8 67 8 21 7 77 7 36 18 0 9 47 8 97 8 50 8 05 7 62 7 22 19 0 9 28 8 79 8 33 7 90 7 48 7 09 20 0 9 09 8 62 8 17 7 75 7 35 6 96 21 0 8 92 8 46 8 02 7 61 7 21 6 84 22 0 8 74 8 30 7 87 7 47 7...

Страница 23: ...83 24 83 630 8 84 10 841 00 97 29 02 737 2 98 29 982 85 82 24 54 623 2 83 09 830 87 96 28 72 729 6 97 27 972 72 81 24 24 615 6 82 07 820 73 95 28 43 722 0 96 26 962 59 80 23 94 608 0 81 06 810 60 94 2...

Страница 24: ...24 APERA INSTRUMENTS LLC Address 6656 Busch Blvd Columbus Ohio 43229 Tel 1 614 285 3080 Email info aperainst com Website https aperainst com...

Отзывы: