Preparation Before First Use 

 

 

 

 

 

 

 

 

2.1 

Pull  out  the  battery  insulation  slip; 

loosen the probe cap ring, then pull out 

the probe cap.  

2.2 

Prepare  a  bucket  of  clean  water  (tap 

water  is  ok.  Ideally  use  RO  water). 

Rinse  the  probe  in  it  and  shake  off 

excess water. 

 

2.3 

Calibrate the tester at pH 7 and pH 4 before first use. See below for calibration tutorial. 

 

pH Calibration 

3.1 

Power  on  the  pen  and  remove the  probe  cap. 

Always  calibrate 

7.00 pH first. 

3.2 

Rinse the probe with clean water and shake off excess water, then 

submerge it in the 7.00 pH standard buffer; 

make a quick stir in 

the solution

 and hold still. 

 

3.3 

Hold 

  until  screen  turns  green.  The  pen  starts  automatic 

calibration.  Wait  for  "Good"  to  show  up  (in  10-15  seconds), 

indicating the calibration is completed, then the pen returns to 

measurement mode. 

 

 

Short press any key while calibrating (in green screen) to cancel 

calibration and return to measurement.

  

 

Содержание 28APE190350/AI31020

Страница 1: ...GroStar GS2 Premium Soil pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...for the horticultural market Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you...

Страница 3: ...the backlight Start calibration In calibration mode press to cancel calibration Switch between F and C Swiss LabSen Spear Soil pH Probe Backed by Apera s proprietary sensor technology from Switzerland...

Страница 4: ...H Calibration 3 1 Power on the pen and remove the probe cap Always calibrate 7 00 pH first 3 2 Rinse the probe with clean water and shake off excess water then submerge it in the 7 00 pH standard buff...

Страница 5: ...f a discrepancy is found then it s time to calibrate again 3 6 If the calibration fails the screen will turn red For details see Section 13 Troubleshooting Guide Always perform at least a 2 point pH c...

Страница 6: ...the white PTFE junction ring using the cleaning brush Avoid brushing the spear glass tip After cleaning shake off excess water For details of probe cleaning refer to Section 8 5 6 Repeat Step 5 2 to...

Страница 7: ...cap Rinse the probe in the water bucket and shake dry 6 8 Submerge the probe into the soil solution shake for a few seconds and wait for the reading to fully stabilize Then record the reading 6 9 Aft...

Страница 8: ...robe in Apera s cleaning solution or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for 1 hour Rinse it off then r...

Страница 9: ...w probe make sure to align the connector s position properly and screw on the probe ring pH probes don t last forever Every pH probe will eventually age and fail even if you don t use it that often Th...

Страница 10: ...fic instrument manufacturers Contaminated probe Clean the probe with Apera s cleaning solution or detergent water Aged probe Replace the probe Dried out probe Soak the probe in the 3M KCl soaking solu...

Страница 11: ...or consistent Contaminated probe After each measurement make sure to use a soft brush with water to clean off all the dirt on the probe especially the white PTFE junction ring Avoid wiping or rubbing...

Страница 12: ...10 pH 10 01 solution sold separately Unit pH F C Power supply 4 AAA alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrumen...

Страница 13: ...receipt for Product showing model number payment and date of purchase This warranty is non transferable and terminates if the original purchaser consumer sells or transfers the Product a third party...

Страница 14: ...horized distributor or reseller How Do You Get Service To begin a warranty claim you must return the Product to the point of purchase with valid proof of purchase Limitation of Liability Acknowledgmen...

Страница 15: ...MAINING PROVISIONS OF THE WARRANTY SHALL REMAIN IN FULL FORCE AND EFFECT Governing Law Authority This warranty is governed by the laws of the state of country where Product is purchased without regard...

Отзывы: