background image

135-04-229-019-EH-0314.pdf

Figure 3

2˚ Max

Slot

Dust Wiper Cover

Mounting

1. To maximize wire rope life, align transducer with mov ing el e ment so that wire rope exits parallel to axis of wire rope exit 

hole within 2° (see Figure 3).

2.  Use Table A or B to determine wire rope exit location relative to trans duc er mount ing holes.

3. Mount unit with two #8 or two M4 (or smaller) pan head ma chine screws.  On units with ranges of 60" (1500 mm) or 80" 

(2000 mm), place spacer blocks under mount ing pods as shown in Figure 2. 

 

Torque 8-32 screws to 

8 lb-in

 max i mum.  

Torque M4 screws to 

0.80 N-m

 maximum.

Rotation of Wire Rope Exit

To rotate wire rope exit location, loosen four screws (item 1, Fig ures 1 & 2) which retain mech a nism to mounting hous ing.  

Note that align ment marks on mech a nism oc cur at 90° in ter vals.  Rotate hous ing to de sired po si tion and torque screws to 

50 oz-in (0.35 N-m)

 maximum.

Replacement Wire Rope Wipers

The JX transducer contains wire rope dust wipers which are lo cat ed just beneath the wire rope exit location.  Periodic 

 

e

c

a

l

e

r

ment of the dust wipers may be nec es sary depending upon the amount of dust in the op 

 t

r

e ing en vi ron ment.  

To access the dust wipers, remove the Dust Wiper Cov er by plac ing a small blade screw driver into the slot as

shown in Figure 3 and pry upward while squeezing the ends of the cover.  The wip ers may be removed with 

tweezers and cleaned or re placed.  The replacement wiper kit which con tains six pair of prelubricated wip ers is 

part num ber 10198.

  To install replacement wipers, place a wiper into wiper open ing on both sides of the wire 

rope.  Press Dust Wiper Cover into place until detented.

*

Mounting screws must accommodate this distance. 

Ranges to 50" (1250 mm)

Ranges to 80" (2000 mm)

Figure 1

Figure 2

*

Mounting screws must ac 

 

o

m

 

m

o

c

date this distance. 

Table A

Option           Output Type                Input Voltage          Output Stage              Waveform                 Cable Wiring                 Optional Connector 

    1       5 VDC TTL Two Channel       +5 VDC ±5%

             Current Sinking

     2      5 V TTL Current Sinking        +5 VDC ±5%

             Differential Line Drive

     3      5 VDC TTL Push Pull            +5 VDC ±5%

             Differential Line Drive

     4      8 to 28 V Current Sinking      +8 to 28 VDC 

             Differential Line Drive

Electrical Output Option

Viewed from solder side 

of mating connector

Note:  With output option 

1 channels A0 and B0 are 

not connected.

Range

Designator

Dim A

(in)     (mm)

Nominal Cable

Tension

(oz)         (N)

10

.36     9.1

16          4.4

15, 30

.50    12.7

14          3.9

20, 40

.66    16.7  11          3.1

25, 50

.82    20.8

 8           2.2

Table B

Range

Designator

Dim A

(in)     (mm)

Nominal Cable

Tension

(oz)         (N)

           60

.98      24.9

10          2.8

80

1.28    32.5

8.6         2.4

JX-EP Series Installation Guide

     5      8 to 28 V Push Pull               +8 to 28 VDC 

             Differential Line Drive

1

JX-EP_1630-21469-0019-E-0314