6 Adjusting and Calibrating
C78IB007EN-B
19
Fig. 15: Methods list – example: long press on a method
TIP: The <Home> button does not save your set-
tings! In case you tap <Home> instead of <OK>,
ViscoQC discards your input.
5.3 Spindle, Spindle Guard,
DIN Adapter
5.3.1 Connecting and Removing a Spindle
Fig. 16: Connecting/removing spindles
ViscoQC features a magnetic coupling. To connect
a spindle, take the spindle shaft beneath the cou-
pling and approach the coupling on the measuring
head from below. Rotate the spindle slightly and
wait until it clicks into place.
To remove a spindle, take the spindle shaft beneath
the coupling and incline the spindle sideways. Then
pull the spindle downwards.
5.3.2 Connecting and Removing a Spindle
Guard or DIN Adapter
The spindle guard is fixed to the rear of the measur-
ing flange with an encoded screw showing the iden-
tification of the guard (
L
or
R
).
If using DIN measuring systems (concentric cylin-
ders CC or double-gap DG) a DIN Adapter is fixed
to the same thread. The fixing screw of a DIN Adapt-
er is marked
D
.
ViscoQC can automatically detect the type of fixing
screw (
1
Removing the fixing screw from the holder
2
Inserting the fixing screw into the holder
3
Mounting a spindle guard / DIN adapter
4
Removing a spindle guard / DIN adapter
Fig. 17: Fixing screw for spindle guard / DIN adapter
The fixing screw is positioned in a threaded hole on
the holder of the guard/DIN adapter. The screw can-
not drop out accidentally due to the hole’s thread. To
remove the fixing screw from the holder you un-
screw it.
Align the holder of the guard or DIN adapter with the
rear of the measuring flange and fasten the screw
(right thread). If you loosen the screw, hold the spin-
dle guard/DIN adapter, so that it does not drop.
6 Adjusting and Calibrating
Adjusting means changing some hardware or soft-
ware factors on the instrument that will have an in-
fluence on future measurement results. Calibrating
means verifying that the instrument measures ac-
cording to specifications: You perform a check mea-
surement with a certified reference substance.
6.1 Calibrating
Verify that ViscoQC measures correctly at least
once a month. Measure a certified reference sub-
stance. Determine how much the measuring result
may deviate from the reference value. Then check