10 Defining General Settings
46
C76IB003EN-E
•
Define mandatory data fields. The measurement can only be started after
filling in the mandatory data fields. If mandatory data fields are not completed
in the sample list, these fields will be prompted again for completion after the
start of the measurement.
To set sample list mode and automatic sample name
1. Tap <Menu> and select
"Setup > Measuring System Settings > Sample
List > Sample List / Sample Settings"
.
2. Select the "Sample List Mode".
•
If no sample changer or a sample changer without magazine is connected,
you can select "No Sample List" or "Sample List".
The "No Sample List" mode is set by default. On the main screen, the button
<Quick Settings> is displayed.
If the "Sample List" mode is selected, the button <Sample List> is displayed
instead of the <Quick Settings> button on the main screen.
•
If a sample changer with magazine is connected, you can select "Simple"
mode or "User-defined" mode. For details, see the corresponding Xsample
instruction manual.
3. Define the "Automatic Sample Name" ("Sample Name Prefix" and "Sample
Name Postfix").
4. Tap <OK> to save the changes.
To set user-defined data fields
1. Tap <Menu> and select
"Setup > Measuring System Settings > Sample
List > User-defined Data Fields"
.
2. Activate/deactivate the three "User-defined Data Fields", enter their names
and tap <OK> to save the changes.
To set sample list warnings
The sample list warnings are only generated if you prepare a sample list prior to
any measurement.
1. Tap <Menu> and select
"Setup > Measuring System Settings > Sample
List > Sample List Warnings"
.
2. Use the two check boxes to activate/deactivate the "Sample List Warnings":
•
Warn if sample list contains a sample without a name.
•
Warn if sample list contains methods with different measurement
temperatures.
3. Tap <OK> to save the changes.