9.4
Selecting the Measurement Mode
DMA 35 provides three different measurement modes - "Precise", "Fast" and
"Manual". According to the measurement mode selected, different stability criteria
have to be fulfilled before the result is stored. The stability criterion is always related
to the temperature:
•
When the measurement mode "Precise" is selected, the result is stored as soon
as the measured temperature value stays within 0.2 K for 10 seconds. This
measurement mode delivers the most accurate results, but may take a longer
time in case the sample temperature differs greatly from the ambient
temperature.
•
When the measurement mode "Fast" is selected, the result is stored as soon as
the measured temperature value stays within 0.4 K for 10 seconds. This
measurement mode delivers quicker results than the "Precise" mode, but as the
density is highly temperature dependent, the measured result is not that
accurate.
•
The measurement mode "Manual" enables you to decide yourself when to store
your measurement result. The result is stored immediately after pressing the
<Store> key on the backside of the instrument.
1.
Press <Menu> and select
"Measurement Mode"
.
2.
Select the preferred setting (Precise, Fast, Manual) and press <Save>.
9.5
Using the RFID Function (Only for Versions with RFID Interface)
The product versions DMA 35 Tag&Log, DMA 35 Ex and DMA 35 Ex Petrol are
equipped with an RFID interface, enabling you to uniquely link an RFID tag to a
method and/or sample ID. The RFID tag can either be allocated to a sample ID and/
or method on the instrument or can be programmed accordingly. By reading the RFID
tag that has been uniquely allocated to a method and/or sample ID or programmed
accordingly, when the instrument displays the main screen, this method and/or
sample ID is automatically used for the following measurements.
9.5.1 Preconditions for Using RFID
The reading range of the RFID receiver integrated in DMA 35 is about 2 cm. The RFID
tag has to be situated within this area to enable data exchange with DMA 35.
TIP
The larger the diameter of a tag, the larger is the reading range.
Содержание 84138
Страница 2: ......
Страница 3: ...C96IB01 3 EN DMA 35 Portable Density Meter From Firmware Version V1 79 Instruction Manual ...
Страница 72: ...72 C96IB01 EN Appendix C CE Declarations of Conformity ...
Страница 73: ...C96IB01 73 EN ...
Страница 74: ...74 C96IB01 EN Appendix D EC Type Examination Certificate ...
Страница 75: ...C96IB01 75 EN ...
Страница 76: ...76 C96IB01 EN Appendix E RFID Module Test Reports ...
Страница 77: ...C96IB01 77 EN ...
Страница 81: ...C96IB01 81 DE DMA 35 Tragbares Dichtemessgerät Ab Firmwareversion V1 79 Betriebsanleitung ...
Страница 153: ...C96IB01 153 DE Anhang C CE Konformitätserklärung ...
Страница 154: ...154 C96IB01 DE ...
Страница 155: ...C96IB01 155 DE Anhang D EG Baumusterprüfbescheinung ...
Страница 156: ...156 C96IB01 DE ...
Страница 157: ...C96IB01 157 DE Anhang E RFID Modul Testreports ...
Страница 158: ...158 C96IB01 DE ...