6.6.
Hydraulic saliva ejecto
r
The hydraulic saliva ejector starts running when the tube is removed from
the support.
Cleaning after each use.
Aspirate about ½ litre of STER 3 PLUS (CEFLA s.c.) diluted in a 6% solution
(equivalent to 60 ml of product in 1 litre of water).
Cleaning the saliva ejector filter
This operation must be carried out at the end of each work day.
WARNING!
Put on gloves before attempting to perform this operation!
• Aspirate about ½ litre of STER 3 PLUS diluted in a 6% solution (equivalent
to 60 ml of product in 1 litre of water).
• In order to prevent possible dripping of liquids and secretions from the
filter (
b
) to be extracted, aspirate only air for about 5 seconds.
• Take off the cap (
a
) by turning and twisting at the same time.
• Remove the filter (
b
).
• Clean/replace the filter (code 97290060).
• Put the filter and cap back in place.
Routine maintenance
Lubricate the o-rings (
c
) with
S1 – Protection for o-rings lubricant
.
Содержание SMART TOUCH R7
Страница 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Страница 2: ......
Страница 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Страница 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Страница 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Страница 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Страница 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Страница 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Страница 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Страница 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Страница 90: ......