34
34
EN
R7
- OPERATING INSTRUCTION
5.2.2.
"Push-pedal” foot control
Description of the parts
1
Handle
2
Control pedal
3
Dental chair movements
4
Chip-air/patient rinsing position control.
5
Water Clean System/Automatic dental chair return control.
6
Spray operation LED
7
Battery charge LED (wireless version only).
Control pedal ( 2 )
Operation:
• Remove the instrument
• Push the foot pedal to start the instrument (
a
)
• Adjust the rpm/power of the instrument with the control pedal:
- to right: to increase
- to left: to decrease
NOTE
: the control pedal adjusts the speed/power of the instrument
from the minimum to the maximum value set from the instrument board.
• To stop the instrument, simply release the control pedal (
a
).
NOTE
: with the spray active, at the end of the operation a blast
of air is automatically activated to remove any residual drops of liquid in
the spray ducts.
WARNING!
Instrument spray is activated and deactivated by pressing the but-
tons ( 4 ) or ( 5 ).
A beep sounds to signal the operating status has been changed.
When the LED ( 6 ) is on, it indicates operation with spray.
Left-hand button operation ( 4 ).
•
Key held down (at least 2 seconds) with the instrument removed:
Chip-air operation: delivers air to the turbine or micromotor.
Air is delivered by pressing the button. Air is no longer blown when the
button is released.
• Key held down (at least 2 seconds) with the instruments in place:
"Rinse position” (PR) program activated.
NOTE
: Press the key again to bring the chair back to the work
position.
Right-hand button operation ( 5 )
•
Key held down (at least 2 seconds) with the instrument removed:
Water Clean System operation: running water is sent to the instruments
such as the turbine, micromotor and scaler to flush the spray ducts.
Water is delivered by pressing the button. Water is no longer delivered
when the button is released and air is automatically blown to eliminate
any drops of liquid remaining in the spray ducts.
•
Key held down (at least 2 seconds) with the instruments in place:
"Dental chair automatic return” program activated.
WIRLESS version.
This foot control can also be supplied in wireless version (see Paragraph
5.2.4).
Protection against liquid penetration.
The foot control is protected against liquid penetration. Degree of protection: IPX1.
Cleaning.
Clean the foot control using a suitable product (see Paragraph 1.4).
NOTE
: If the foot control slips on the floor, dust the slip-proof rubber found under the base with a dry cloth
.
Содержание SMART TOUCH R7
Страница 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Страница 2: ......
Страница 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Страница 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Страница 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Страница 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Страница 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Страница 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Страница 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Страница 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Страница 90: ......