3-6 Directivity and Test Port Match Verification
Chapter 3 — Performance Verification
3-14
PN: 10410-00730 Rev. D
MN469xC Series Multiport Test Set MM
78.
Position Mkr 2 to the bottom of the trough (or to the top of the peak if the ripple has the greatest position
peak.)
79.
Sum the magnitude values of Mkr 1 and Mkr 3 at the peaks (or troughs) and divide the result by two.
This is the average value of the two peaks (or troughs) Refer to the example formula below:
Average Value = (Mkr 1
Mkr 2)/2 = (
15.9634 dB)
(
15.641 dB)/2 =
15.8022 dB
80.
Calculate the peak-to-peak ripple value (absolute difference of the Mkr 2 value and the Average Value) as
follows:
dB
p-p
= |Mkr 2 value|
|Average Value| = 17.452 dB
15.8022 dB = 1.6498 dB
81.
Use a RF measurement chart in
to find the corresponding return loss value of the
peak-to-peak ripple value.
•
Example: The corresponding Return Loss value of 1.6498 dB
p-p
is approximately 20 dB
Figure 3-6.
Log Mag Display