Performance Specifications
Performance specifications are provided in Table 1-1.
1 - 5
Chapter 1 General Information
Specifications are valid when the unit is calibrated at ambient tem-
perature after a 5 minute warmup.
Description
Value
Frequency Range:
Site Master S112
Site Master S113
5 to 1000 MHz
5 to 1200 MHz
Frequency Accuracy (CW Mode)
75 parts per million @25
°
C*
Frequency Resolution
10 kHz
SWR:
Range
Resolution
1.00 to 65.00
0.01
Return Loss:
Range
Resolution
0.0 to 54.00 dB
0.01 dB
Cable Insertion Loss:
Range
Resolution
0.0 to 20.00 dB
0.01 dB
**Distance-To-Fault (DTF):
Range
Resolution (in meters)
(Rectangular Windowing)
0 to (Resoution x 129)
( .
)(
)
15
10
8
´
V
Frequency
p
D
Where
V
p
is the cable’s relative
propagation velocity.
Table 1-1.
Performance Specifications (1 of 2)