4-9 Single Sideband Phase Noise Test
Performance Verification
4-14
PN: 10370-10386 Rev. B
MG362x1A MM
2.
E5052B and E5053A setup:
a.
Press the green
Preset
button, then touch
Factory
.
b.
Press
Measure/View
, then touch
Phase Noise
.
c.
Press
Window Max
.
d.
Press
Start/Center
, then touch
10 Hz (enter 1 Hz if phase noise options are installed)
.
e.
Press
Stop/Span
, then touch
1 MHz.
f.
For standard units and units with low phase noise options, press
Marker
, then:
Touch
Marker 1
, enter
10 then press X1
to obtain 10 Hz.
Touch
Marker 2
, enter
100 then press X1
to obtain 100 Hz.
Touch
Marker 3
, enter
1 then press k/M
to obtain 1 kHz.
Touch
Marker 4
, enter
10 then press k/M
to obtain 10 kHz.
Touch
Marker 5
, enter
100 then press k/m
to obtain 100 kHz.
Touch
Marker 6
, enter
1 then press M/u
to obtain 1 MHz.
Figure 4-4.
Equipment Setup for Single Sideband Phase Noise Test with E5052B and E5053A
MG362x1A
Splitter
E5052B Signal Source Analyzer with
E5053A Down Converter
RF Output
50 Ω
RF On/Off
Unleveled
Output: +29 dBm max
Reverse: +27 dBm, 0 VDC
FREQ
LEVEL
USER
SELECT
0
1
2
3
4
5
6
7
8
9
MG36221A
Signal Generator
20 GHz
Содержание Rubidium MG362 1A Series
Страница 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Страница 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Страница 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Страница 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Страница 164: ...4 26 Pulse Modulation Tests with the 86100C Oscilloscope Performance Verification 4 98 PN 10370 10386 Rev B MG362x1A MM ...
Страница 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Страница 293: ......