6.2 Devices Required for Performance Tests
6-3
6.2 Devices Required for Performance Tests
Before starting performance tests, warm up the MU181800A/B and the
measuring instruments for at least 30 minutes. Table 6.2-1 shows the
devices required for performance tests.
Table 6.2-1 Devices required for MU181800A/B performance test
Device
Required Performance
Synthesizer
Generating frequency:
0.1 to 12.5 GHz (MU181800A)
0.1 to 14.0 GHz (MU181800B)
0.1 to 14.1 GHz (MU181800B-005)
Output amplitude:
0.4 to 2.0 Vp-p
Sampling oscilloscope
Electrical interface:
50 GHz or more band
Trigger signal source
(M MU181020A, or
1/64 divider supporting 14
GHz)
Generates trigger to observe the sampling oscilloscope waveform
Note:
Before starting the performance tests, warm up the device under
test and the measuring instruments for at least 30 minutes and
wait until they become sufficiently stabilized, unless otherwise
specified. Additional conditions are required for maximum
measurement accuracy: measurements must be performed at room
temperature, fluctuations of AC power supply voltage must be
small, and noise, vibration, dust, and humidity must be
insignificant.
Содержание MU181800A
Страница 10: ...x ...
Страница 14: ...IV ...
Страница 20: ...Chapter 1 Overview 1 6 ...
Страница 24: ...Chapter 2 Preparation before Use 2 4 ...
Страница 30: ...Chapter 3 Panel Layout and Connectors 3 6 ...
Страница 40: ...Chapter 6 Performance Test 6 6 ...
Страница 47: ...Appendix App I Appendix A Performance Test Result Sheet A 1 ...
Страница 48: ...Appendix App II ...
Страница 50: ...Appendix A Performance Test Result Sheet A 2 ...