Power Connection
3-2
Battery Operation
•
A fully charged battery has a 94 watt-hour capacity.
•
When the battery is charged to 100% the four status LEDs will blink in unison on and off when the
battery indicator is switched on. If the external AC/DC power supply is connected when there is no
battery, then the four LEDs will blink randomly.
•
Removing the charger or battery while operating does not interrupt the power supplies or the USB
communications.
Warning
When supplying power to the equipment, connect the accessory power supply to a 3-pin grounded
power outlet connected in turn to local AC Mains. If a grounded 3-pin outlet is not available, use a
conversion adapter and ground the green wire, or connect the equipment frame to a suitable ground.
If power is supplied without grounding the equipment, there is a risk of receiving a severe or fatal
shock.
Note
For maximum ESD protection, proper grounding is required. Connect the accessory power supply to
a three pin grounded power outlet and use the ground receptacle on the front panel of the unit to
ground the chassis.
Note
With the battery installed the unit is active as long as there is still charge on the battery. There is no
power ON/OFF switch.
Warning
To insure proper operation, only use the 40-204-R power supply (15V, 80W, with a right angle
connector) with the MN25131A Multi-Function Extender.