Chapter 3 — Performance Verification
3-7 Spurious Signals Tests
MG3702xA MM
PN: 10370-10372 Rev. B
3-9
2.
Set up the Spectrum Analyzer as follows:
a.
Press the
AMPLITUDE
key.
b.
Press
REF LVL
and enter the current power level setting (L0) of the MG3702xA.
c.
Press the
FREQUENCY
key.
d.
Press
Start Freq
.
e.
Enter the current spectrum analyzer start frequency from the test record.
f.
Press
Stop Freq
.
g.
Enter the current spectrum analyzer stop frequency from the test record.
h.
Press the
BW
key.
i.
Press
Res BW to MAN
.
j.
Enter the current spectrum analyzer RBW value from the test record.
3.
Press the
PEAK SEARCH
key, and then select
MARKER DELTA
.
4.
Set up the Spectrum Analyzer as follows:
a.
Press the
FREQUENCY
key.
b.
Press
Start Freq
.
c.
Enter the next spectrum analyzer start frequency from the test record.
d.
Press
Stop Freq
.
e.
Enter the next spectrum analyzer stop frequency from the test record.
f.
Press the
BW
key.
g.
Press
Res BW to MAN
.
h.
Enter the next Spectrum Analyzer RBW value from the test record.
5.
Press
PEAK SEARCH
.
6.
Record the reading from the Spectrum Analyzer Delta MKR reading in
Non-Harmonic Signals Test (Part 1)”
, Part 1.
7.
for same specified MG3702xA frequency listed in the test record, part
1.
8.
Repeat
through
for same each MG3702xA frequency groups listed in the test record,
part 1.
Non-Harmonic Test Procedure: Part 2 (with option 004 installed)
For the following test, refer to the test record table in
“Section 3-7 Non-Harmonic Signals
“Non-Harmonic Test Procedure: Part 2 (with option 004 installed)”
.
1.
Set up the MG3702xA as follows:
a.
Press Edit Frequency to open the current frequency parameter for editing.
b.
Set F0 to the appropriate test frequency (based on model and options) in the test record, part
2.
c.
Press
Edit Level
to open the current power level parameter for editing.
d.
Set L0 to +10 dBm or to the maximum specified power level, whichever is less. Refer to
“MG3702xA Technical Data Sheet”
for the maximum specified power levels.
For the electronic version, refer to the MG3702xA Technical Data Sheet, PN: 11410-00429
on the Anritsu Internet site,
http://www.anritsu.com.
2.
Set up the spectrum analyzer as follows:
a.
Press the
AMPLITUDE
key.
Содержание MG3702xA Series
Страница 4: ......
Страница 5: ......
Страница 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Страница 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Страница 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Страница 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 144: ...6 13 RF Components Chapter 6 Removal and Replacement Procedures 6 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 178: ...B 1 MG3702xA Technical Data Sheet Appendix B Performance Specifications B 2 PN 10370 10372 Rev B MG3702xA MM ...
Страница 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Страница 183: ......