7-2
Master-Slave
Operation
Master-slave operation consists of connecting any two MG369XB in-
struments together and configuring them so that they produce CW
and synchronized, swept output signals at an operator-selectable fre-
quency offset. One instrument (the Master) controls the other (the
Slave) via interface cables between their rear panel
AUX I/O
and
SE-
RIAL I/O
connectors. The two units are phase-locked together by con-
necting them to the same 10 MHz reference time base.
Connecting
the
Instruments
Connect the two instruments, shown in Figure 7-1,
as follows:
Step 1.
Connect the 3-port AUX I/O cable end la-
beled “MASTER” to the rear panel
AUX
I/O
connector on the Master instrument.
Connect the AUX I/O cable labeled
“SLAVE” to the rear panel
AUX I/O
con-
nector on the Slave instrument.
Step 2.
Connect the ends of the flat interface ca-
ble to the rear panel
Serial I/O
connectors
on the Master and Slave instruments.
Step 3.
Connect one end of a coaxial cable to the
rear panel
10 MHz REF OUT
connector on
the Master instrument. Connect the other
end to the rear panel
10 MHz REF IN
con-
nector on the Slave instrument.
7-4
MG369XB OM
Master-Slave Operation
Use With Other Instruments
R F O u t p u t
M G 3 6 9 X B ( M a s t e r )
M G 3 6 9 X B ( S l a v e )
T O D U T
T O D U T
1 0 M H z
R E F O U T
1 0 M H z
R E F I N
A U X
I / O
A U X
I / O
R F O u t p u t
S E R I A L
I / O
S E R I A L
I / O
T O
5 6 1 0 0 A
S N A
Figure 7-1.
MG369XB Configuration for Master-Slave Operation
NOTES
When connecting two instruments
together for Master-Slave opera-
tions, always use an Anritsu Mas-
ter-Slave interface cable set, Part
No. ND36329.
If a Model 56100A Scalar Network
Analyzer is being used with the
master-slave configuration, (1) con-
nect the AUX I/O cable end labeled
“SNA” to the rear panel
AUX I/O
connector on the 56100A SNA and
(2) connect a dedicated system bus
cable (P/N 2100-1) between the
Master instrument rear panel
IEEE-488 GPIB
connector and the
56100A SNA rear panel
DEDI-
CATED GPIB
connector.
Содержание MG369 Series
Страница 3: ......
Страница 4: ......
Страница 14: ...1 2 MG369XB OM Figure 1 1 Series MG369XB Synthesized Signal Generator ...
Страница 20: ......
Страница 22: ......
Страница 24: ...Preparation For Use Installation 2 4 MG369XB OM Figure 2 1 MG369XB Outline Dimensions in millimeters ...
Страница 38: ......
Страница 154: ......
Страница 155: ...Table of Contents 4 1 Introduction 4 3 4 2 Menu Map Description 4 3 Chapter 4 Local Operation Menu Maps ...
Страница 156: ......
Страница 158: ......
Страница 174: ......
Страница 214: ......
Страница 216: ......
Страница 230: ......
Страница 232: ......
Страница 260: ......
Страница 262: ......
Страница 268: ......