3-18
Scan Modulation
(Option 20)
The internal scan modulator, added by Option 20, lets you amplitude
modulate output signals from 2 to 18 GHz at modulation depths up to
60 dB. Scan modulation is accomplished using a modulating signal
from an external source and is in addition to the normal amplitude
modulation described in Section 3-12.
To provide amplitude modulation of the output sig-
nal using the internal scan modulator, first set up
the external modulating signal generator and con-
nect it to the MG369XB’s rear panel
SCAN MOD IN
connector.
Next, press
MODULATION
, then
Scan
to go to the
Scan Status menu (below).
Press
On/Off
to turn scan modulation on and off. The
Scan status display and the Scan modulation status
area will reflect your selection as On or Off.
Control the scan modulation depth by varying the
level of the external modulating signal. The scan
modulator has a fixed sensitivity of –10 dB/V. An ex-
ternal modulating signal level of 6 volts produces
the maximum modulation depth of –60 dB.
MG369XB OM
3-119/3-122
Local (Front Panel) Operation
Scan Modulation (Option 20)
Содержание MG369 Series
Страница 3: ......
Страница 4: ......
Страница 14: ...1 2 MG369XB OM Figure 1 1 Series MG369XB Synthesized Signal Generator ...
Страница 20: ......
Страница 22: ......
Страница 24: ...Preparation For Use Installation 2 4 MG369XB OM Figure 2 1 MG369XB Outline Dimensions in millimeters ...
Страница 38: ......
Страница 154: ......
Страница 155: ...Table of Contents 4 1 Introduction 4 3 4 2 Menu Map Description 4 3 Chapter 4 Local Operation Menu Maps ...
Страница 156: ......
Страница 158: ......
Страница 174: ......
Страница 214: ......
Страница 216: ......
Страница 230: ......
Страница 232: ......
Страница 260: ......
Страница 262: ......
Страница 268: ......