Press
On / Off
to turn AM on and off. The External
AM status display will reflect your selection as On
or Off; the AM modulation status area will reflect
your selection as External On or Off.
Press
Edit Sensitivity
to open the AM Sensitivity pa-
rameter, then edit the current value using the cur-
sor control keys, rotary data knob, or enter a new
value using the keypad and the appropriate termi-
nator key (kHz/ms/STEPS for Linear; MHz/ms/dB
for Log). The AM Sensitivity range is 0 %/V to
100 %/V in Linear and 0 dB/V to 25 dB/V in Log. To
close the open AM Sensitivity parameter, press
Edit
Sensitivity
or make another menu selection.
Press
Internal / External
to select the source of the
modulating signal. Internal selects the modulating
signal from the internal AM generator; external se-
lects the modulating signal from an external source.
The AM status display will reflect your selection.
Press
Log / Linear
to select the AM operating mode.
When Internal AM is active, the AM Depth display
will reflect your selection as XX dB (Log) or XX %
(Linear). When External AM is active, the AM Sen-
sitivity display will reflect your selection as XX dB/V
(Log) or XX %/V (Linear).
Press
< Previous
to return to the AM Status menu
display.
Frequency
Modulation
Operating
Modes
The signal generator has four FM operating modes:
Locked, Locked Low-Noise, Unlocked Narrow, and
Unlocked Wide. In the Locked and Locked
Low-Noise FM modes, frequency modulation of the
output signal is accomplished by summing the mod-
ulating signal into the FM control path of the YIG
phase-lock loop.
In Locked FM mode, the maximum FM deviation is
the lesser of
±
10 MHz or the rate times 300 for
1 kHz to 8 MHz rates. In Locked Low-Noise FM
mode, the maximum FM deviation is the lesser of
±
10 MHz or the rate times three for 50 kHz to
8 MHz rates.
3-96
MG369XB OM
Signal Modulation
Local (Front Panel) Operation
ERR
This error message is displayed
when the external AM modulating
signal exceeds the input voltage
range. The message “
Reduce AM In-
put Level
” also appears at the bot-
tom of the AM status display. AM is
turned off until the modulating sig-
nal is within the input voltage
range.
Содержание MG369 Series
Страница 3: ......
Страница 4: ......
Страница 14: ...1 2 MG369XB OM Figure 1 1 Series MG369XB Synthesized Signal Generator ...
Страница 20: ......
Страница 22: ......
Страница 24: ...Preparation For Use Installation 2 4 MG369XB OM Figure 2 1 MG369XB Outline Dimensions in millimeters ...
Страница 38: ......
Страница 154: ......
Страница 155: ...Table of Contents 4 1 Introduction 4 3 4 2 Menu Map Description 4 3 Chapter 4 Local Operation Menu Maps ...
Страница 156: ......
Страница 158: ......
Страница 174: ......
Страница 214: ......
Страница 216: ......
Страница 230: ......
Страница 232: ......
Страница 260: ......
Страница 262: ......
Страница 268: ......