Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-27
DFM <value> <unit>
Description:
Opens the delta frequency parameter.
Permits setting the Delta Frequency parameter. The frequency setting is determined by
the frequency range of the particular MG369xC model. The commands, DFF and DLF,
are the same as DFM.
Functional
Category:
PARAMETER ENTRY
Value:
Frequency range of the MG369xC model
Units:
GH, MH, KH, HZ
Example:
F1 3 GH; DFM 100 MH; DF1; SSP; AUT
Sets auto trigger step sweep F1-
F 3GHz-50MHz to 3GHz+50MHz.
Output Command:
ODF
returns the DF frequency value (in MHz) to the controller.
Front Panel Key:
N/A
DG
Description:
Degree data terminator.
Functional
Category:
DATA TERMINATOR
Front Panel Key:
N/A
DL1
Description:
Selects external detector leveling of output power.
Selects external leveling of the output power using a signal from an external detector
connected to the
EXT ALC IN
connector. Deselects the internal leveling mode or external
leveling mode using a signal from an external power meter, if r previously programmed.
In external power leveling mode, use the parameter entry command EGI to enter a
setting for the Reference Level DAC to control the ALC gain.
Functional
Category:
POWER LEVELING
Example:
DL1; EGI 128 SPS
Selects external leveling of the output power using a signal from an external detector and
sets the reference level DAC to 128.
Related
Commands:
EGI
,
IL1
,
PL1
,
LV0
,
LV1
Front Panel Key:
Level
| ALC Mode | Leveling > | External Detector