Coverage Mapping (Option 431)
6-4 Coverage Mapping
Spectrum Analyzer MG
PN: 10580-00349 Rev. M
6-7
Map the Signal Strength
Coverage Mapping supports RSSI measurement or ACPR measurement during mapping.
ACPR
1.
Press the
Coverage Mapping
submenu key.
2.
Press the
Measurement Setup
submenu key.
3.
Set the
Center Freq
,
RBW
, and
Detection
type. Refer to
Chapter 2, “Spectrum Analyzer”
for additional information.
4.
Press
ACPR
once to select and again to open the setup menu.
a.
Enter the Main and Adjacent Channel Bandwidths.
b.
Enter the Channel Spacing.
c.
Enter Good Passing Criteria and the Poor threshold level.
d.
The main channel power indicator in the bottom part and the data collection
squares will display colors as shown below:
Red Value
(Poor) <
Yellow Value
<
Green Value
(Good)
5.
Press the
Start Data Collection
main menu key. Data will be collected at the time or
distance interval based on the setting in
“Point Distance/Time Setup Menu”
. The color of the squares indicates the power level based on ACPR setup.
You may have to move the instrument around, and for indoor maps, you may need to
tap the touch screen.
6.
Press the
Stop Data Collection
main menu key. Save the collected data as a .kml file,
a tab-delimited text file (
.mtd
) or a
.jpg
file. Refer to
.
Note
The default settings for coverage mapping set the internal input attenuator to 0 dB
(off) with the preamp on and a bandwidth of 1 MHz. These parameters can be
adjusted depending on the strength of the signal being measured. The bandwidth
is 10 kHz for RSSI measurements. For ACPR, it is based on ACPR settings
(channel width and spacing) and the resultant span, as well as if Auto RBW is set.
Overdriving the input of the spectrum analyzer can result in ADC Overdrive or
Saturation errors. Refer to
“Indications of Excessive Signal Level” on page 2-10
necessary, make adjustments using the
Amplitude
menu and
BW
menu.
Once data collection begins, parameters such as amplitude, bandwidth or
measurement setup cannot be changed.
Note
The collected data can be saved in multiple formats.
Warning
Using Anritsu equipment while operating a motor vehicle is dangerous and
could lead to serious accidents. Check the laws and regulations that are in
effect in your area with regard to the use and placement of electronic devices
or fixtures on a moving vehicle.
Содержание BTS Master
Страница 103: ...Spectrum Analyzer 2 35 Trace Menu Spectrum Analyzer MG PN 10580 00349 Rev M 2 91 Figure 2 78 Trace Info...
Страница 246: ...6 6 Coverage Mapping Menu Coverage Mapping Option 431 6 24 PN 10580 00349 Rev M Spectrum Analyzer MG...
Страница 278: ...7 21 Marker Menu AM FM PM Analyzer Option 509 7 32 PN 10580 00349 Rev M Spectrum Analyzer MG...
Страница 282: ...8 2 Bias Tee Menu Bias Tee Option 10 8 4 PN 10580 00349 Rev M Spectrum Analyzer MG...
Страница 295: ......